it to adjust to room temperature (approximately 2 hours) before beginning setup.
• Make sure the device is away from any heating or cooling equipment (e.g.,
forced air vents, radiators, air conditioners).
• The device is not suitable for use in high humidity environments. Make sure that no
water enters the device.
• Make sure that bedding, curtains, or other items are not blocking the filter or vents
of the device.
• Keep pets, pests or children away from the device and avoid small objects being
inhaled or swallowed.
• To avoid explosion, this device must not be used in the presence of flammable
gases (e.g. anesthetics).
• Tobacco smoke may cause tar build-up within the device, leading to the
malfunctioning of the device.
• Air must flow freely around the device for it to work properly.
• Empty the water chamber completely before moving the device.
12.2 Installing the Air Filter and Filter Cap
(1) Attach the air filter to the filter cap, as shown in Fig. 12-1.
Fig. 12-1
(2) Install the filter cap containing the air filter to the device, as shown in Fig. 12-2.
Fig. 12-2
WARNINGS!
• Do not block the gas INTAKE PORT, thereby interfering with the therapy.
• Nebulization or humidification can increase the resistance of breathing system
filters and the operator must monitor the breathing system filter frequently for
increased resistance and blockage to ensure the delivery of the therapeutic
pressure.