18. Cleaning
WARNINGS!
• Cleaning the device and its accessories as recommended in the following
sections is essential to prevent respiratory infections.
• To avoid electric shock, always unplug the device before cleaning.
• Use mild soap that is nontoxic to humans.
• Follow the manufacturer's instructions on cleaning the mask and tubing and on
determining the frequency of cleaning.
• Before cleaning, check that the device is disconnected from the power supply,
whether the power cord is unplugged, and whether the water chamber of the
device has cooled down. Make sure that the heating plate has cooled down to
room temperature, so that you do not get burned.
• Do not open or modify the device. There are no operator serviceable parts inside.
Repairs and service should only be performed by an authorized service agent.
• Avoid the use of any CPAP cleaner or disinfection device that relies on ozone (i.e.
activated oxygen). Device warranty may terminate if the damage is caused by the
use of an ozone cleaner.
• In order to prevent contamination of the device, use only
manufacturer-approved filters on this device conforming to ISO 23328-1:2003 and
ISO 23328-2:2002 standards.
• The device shall not be serviced or maintained while a patient is using it.
CAUTIONS!
• Overheating of the materials could lead to early fatigue of the materials.
• Do not use solutions containing chlorinated lime, chlorine, or aromatic to clean
the device and its accessories. Liquid soap containing moisturizing agents or
antimicrobials should not be used, either. These solutions may harden cleaned
materials or reduce their lifespan.
• Do not clean or dry the device and its accessories when the temperature is above
80°C (176°F ). High temperatures could reduce product life.
• Do not immerse the device in any fluids.
Accessories that
need to be
cleaned
Warm water
(approx 113 to 140°F or 45 to 60°C)
Warm water
(approx 113 to 140°F or 45 to 60°C)
The tap water
5°C to 35°C (41˚F to 95˚F)