Luna
®
G3 APAP / Luna
®
G3 CPAP User Manual
34 / 49
Fig. 18-2
(2) Cleaning the Transfer Box: Dilute Alconox to 1% with water 45°C to 60°C (113°F to
140°F). Soak the transfer box in the detergent for 5 minutes. Clean the transfer box
with a soft bristled brush for one minute. Pay particular attention to all gaps and
cavities. Then rinse the transfer box with running water for 5 minutes. Wipe it dry with
a soft cloth or air dry it away from direct sunlight.
(3) Putting the Transfer Box back: as shown in Fig. 18-3.
Fig. 18-3
CAUTION!
• It is recommended to clean the transfer box once a week.
18.3 Cleaning the Mask and Headgear
For details, refer to the cleaning instructions in the user manual for the mask.
18.4 Cleaning the Enclosure
Wipe the enclosure of the device with a soft, slightly damp cloth.
CAUTIONS!
• The device can only be used after the enclosure is dry, so that no moisture enters
the device.
• It is recommended to clean the enclosure once a week.
18.5 Cleaning the Tubing
(1) Remove the tubing from the device and mask before cleaning.
(2) Hold the cuff of the tubing and gently pull it away from the device as shown in
Fig. 18-4. Or disconnect the power of the heated tubing, then hold the cuff of the
heated tubing and pull it away from the device, as shown in Fig. 18-5.