Luna
®
G3 APAP / Luna
®
G3 CPAP User Manual
6 / 49
6. Specifications
Device Size
Dimensions (L x W x H): 265 mm × 145 mm × 114 mm
Weight: 1.7 kg
Water capacity: To maximum fill line 360 mL
Product Use, Transport and Storage
Operation Transport and Storage
Temperature: 5°C to 35°C (41˚F to 95˚F) -25°C to 70°C (-13˚F to 158˚F)
Humidity: 15% to 93% Non-condensing 15% to 93% Non-condensing
Atmospheric Pressure: 760 to 1060 cmH
2
O 760 to 1060 cmH
2
O
Altitude: Sea level to 2300 m Sea level to 2300 m
Heated Humidifier
Humidifier Settings: Off, Auto, 1 to 5 (95˚F to 154.4˚F / 35°C to 68°C)
Humidifier Output: No less than 15 mg H
2
O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity
Maximum Operating Pressure: 40 cmH
2
O
Pressure Drop with Humidifier: < 0.4 cmH
2
O at 60 LPM flow
Maximum Delivered Gas Temperature: ≤ 43°C
Mode of Operation
Continuous
Work Mode
LG3600: CPAP, AutoCPAP
LG3500: CPAP
SD Card
The SD card can record patient data and fault information
AC Power Consumption
100 V - 240 V , 50 Hz / 60 Hz, 2 A Max
Device offer to Heated Tubing Communications Port
24 V 18 W
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IP22
Pressure Range
4 to 20 cmH
2
O (in 0.5 cmH
2
O increments) ≤ 30 cmH
2
O under single fault conditions.