EasyManuals Logo
Home>BMC>Respiratory Product>REACTHEALTH Luna G3 APAP

BMC REACTHEALTH Luna G3 APAP User Manual

BMC REACTHEALTH Luna G3 APAP
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
Luna
®
G3 APAP / Luna
®
G3 CPAP User Manual
24 / 49
14.3 Setting the Humidity Level
After the device is powered on, turn the Knob to turn on or turn off the heated
humidifier and to adjust the humidity level according to instructions of the Patient
Menu of the device.
There are five humidity levels available, and the number of humidity level will
appear in the Main Interface on the screen of the device. The number 2 next to the
icon indicating the humidity is adjusted to Level 2, as shown in Fig. 14-8. The
temperature of the water in the water chamber maintains a constant set level.
Fig. 14-8
WARNING!
• Do not touch the heater plate of the device when it is working, otherwise you may
get burned. Turn off the heat when the heated humidifier is not in use.
CAUTIONS!
• Generally, the humidity inside the mask is low when the water temperature is low.
The greater the difference between the temperature inside the air tubing and
room temperature is, the more easily condensation occurs inside the tubing.
• If there are only a few condensed water droplets inside the tubing in the morning
after therapy, it means that the humidity level is appropriate; if there is lots of
condensed water droplets inside the tubing and / or mask, it means that the
humidity level is too high and should be set lower; Nasal dryness means that the
humidity level is too low and should be set higher.
15. Using the Cellular Module
The Luna
®
G3 APAP / Luna
®
G3 CPAP with a Cellular Module can wirelessly
communicate with the iCodeConnect cloud platform. iCodeConnect cloud
platform is intended to augment the standard follow-up care of patients diagnosed
with obstructive sleep apnea by displaying usage and therapeutic information that
has been transmitted from the patient's Luna
®
G3 APAP / Luna
®
G3 CPAP therapy
device to the clinician or healthcare professional.
(1) Insert the Cellular Module into the device, and turn on the device. The device
screen displays the Main Screen shown in Fig. 15-1.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC REACTHEALTH Luna G3 APAP and is the answer not in the manual?

BMC REACTHEALTH Luna G3 APAP Specifications

General IconGeneral
BrandBMC
ModelREACTHEALTH Luna G3 APAP
CategoryRespiratory Product
LanguageEnglish