8200-M129-O1 Rev D 205 Installation & Technical 24
SETUP AND CALIBRATION, CONT.
If you press the TARE/ENTER key at the SetUP prompt, you may proceed through to the next
section (up to and including f SPAn ) by pressing the TARE/ENTER key.
If you press the ZERO key, dFLtS? is displayed. This allows all setup parameter values to be
replaced with predetermined defaults.
IMPORTANT! Setup may be interrupted at any time. ALL data previously
entered and finalized with the TARE/ENTER key will be retained in the non-
volatile memory.
Pressing the calibration switch at any prompt will return you to the SEtUP menu. To exit
setup, press the /UP ARROW key with any of the above menu selections displayed or cycle
power at any time (press the ON/OFF key twice).
SEtUP
USA= (Domestic or International)
With SEtUP displayed, press the TARE/ENTER key. The display will change to USA=. Press
the TARE/ENTER key to show the current value. If the setting displayed is acceptable, press
the TARE/ENTER key again to save it. Otherwise, use the /UP ARROW key to toggle to a
new setting and then press the TARE/ENTER key to save it.
USA=1 (Domestic) USA=0 (International)
Trl = no Trl = yes
Cap + 4% to OC Cap + 9 grads to OC
PT printed with tare
Lamp test on power up
NOTE: With the exception of the SEtUP prompt, the prompts displayed for each section
are different if you push the calibration switch instead of pressing the TARE/ENTER key to
proceed through the section.
For example, if you press the calibration switch with the SEtUP displayed, the next prompt
displayed will be A-d.
If you step through the setup prompts by pressing the TARE/ENTER key, the next prompt
displayed will be A-d?.
In addition, at a prompt with the ? displayed, you must press the TARE/ENTER key, then
the /UP ARROW key to select YES and then the TARE/ENTER key again to proceed
with that section.
To skip the section and advance you to the next menu selection, press the TARE/ENTER
key twice.