8200-M129-O1 Rev D 205 Installation & Technical 37
SETUP AND CALIBRATION, CONT.
Sst? Setup Guardian Scale
If a GUARDIAN scale is connected to the indicator, at the SST? prompt press the /UP
ARROW key to select YES.
The tLo= prompt will be displayed. This is the low end of the range of weight that requires
adjusting. For example, if after performing linearization testing, the displayed weight is 5 lbs
low between 10,000 lbs and 15,000 lbs, the value for the tweak low weight is 10000.
Using the /UP ARROW and UNITS/LEFT ARROW keys, input the tweak low weight value
and then press the TARE/ENTER key.
The tHi= prompt will be displayed. This is the high end of the range of weight that requires
adjusting. Using the example above, the value for the tweak high weight is 15000.
Using the /UP ARROW and UNITS/LEFT ARROW keys, input the tweak high weight
value and then press the TARE/ENTER key.
The tUAL= prompt will be displayed. This is the amount of weight that needs to be added
(or subtracted) from the range of weight to bring the displayed weight into tolerance. Again,
using the example above, the Tweak Weight value would be 5.
Using the /UP ARROW and UNITS/LEFT ARROW keys, input the tweak weight value
(pressing the NET key will change the sign) and then press the TARE/ENTER key to save it
and advance to the Sio? menu.