EasyManuals Logo

Cardinal 205 User Manual

Cardinal 205
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
8200-M129-O1 Rev D 205 Installation & Technical 53
SETUP REVIEW
The 205 indicator allows several operational parameters to be reviewed and changed without
breaking the calibration seal. These operational parameters are:
Power Up Zero Reset Enable/Disable
Sleep Mode Feature Enable/Disable
Auto Shutoff Feature Enable/Disable
Serial Input / Output Configuration
Baud Rate
Parity
Number of Data Bits
Number of Stop Bits
Continuous Output Port 1
Continuous Output Format
Continuous Output Port 2
Continuous Output Format
Print Tab Settings
Printer Port Selection
Gross Weight
Tare Weight
Net Weight
Gross Weight Accumulator
Net Weight Accumulator
Carriage Return Line Feed
End Of Print Line Feeds
To enable the Setup Review feature, with the indicator ON:
1. Press the /UP ARROW key. The indicator will respond by showing the FUnCt=
(Function) prompt and alternately flashing off and on (all ON, weighing unit off, then all
OFF, weighing unit ON) the annunciators.
2. Press the ZERO/REVIEW key. The display will change to the prompt for the selection of
power-up zeroing (PUO=).
3. Using the same procedure as described in the Setup and Calibration section of this
manual, make the required changes.
4. Press the /UP ARROW key to return to the previous prompt.
5. To exit Setup Review, press the /UP ARROW key to step through the remaining
prompts OR at anytime, cycle the power (press the ON/OFF key twice).

Questions and Answers: