Note the action time of the cleaning agent.
❯
Protect the unit from contamination when
transporting it from the treatment chair to the
reprocessing location.
10.4 Manual cleaning, intermedi-
ate rinsing, disinfection, final
rinse, drying
A disinfectant or combined cleaning and disinfec-
tant agent is required for manual disinfection. It
must have the following properties:
– certified, possibly virucidal efficacy (DVV/RKI,
VAH or European Standards)
For further information, see: "General informa-
tion".
Cleaning
❯
Place the individual components in a disinfec-
tant bath (non-fixing/aldehyde-free, see "Gen-
eral information") so that all parts are covered.
❯
Comply with the reaction times of the cleaning
agent and disinfectant (see "General informa-
tion").
❯
If you notice any further contamination, brush
all exterior and interior surfaces completely with
a hygienic brush under the surface of the
ready-to-use solution.
Intermediate rinsing
After the action time prescribed by the manufac-
turer:
❯
Rinse off all components under water for at
least 1 minute (temperature < 35°C).
Disinfection
❯
Place individual components in a cleaning and
disinfectant bath so that all parts are covered.
❯
Note the action time for the disinfectant.
Final rinse
After the action time prescribed by the manufac-
turer:
❯
Rinse off all components under water for at
least 1 minute (temperature < 35°C).
Drying
❯
If necessary, re-dry at a clean location using a
hygienic, lint-free cloth.
❯
Blow dry the components with compressed air
in a clean location.
Usage
9000-618-176/30 1812V009 29
EN