Replacement Procedures
8-10 LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5
Air Filter Cleaning
Clean the system’s air filter to ensure that a clogged filter does
not cause the system to overheat and reduce system
performance and reliability. It is recommended the filters be
cleaned every two weeks, but the requirements will vary with
environment.
Tools
• None
Preparation
Shut down the system and disconnect the power cord.
Needed Manpower
1 person, 5 minutes + travel
CAUTION
Be sure to lock the wheels before cleaning the air filters to
aviod injury by any unexpected movement of the system.
DO NOT operate the unit without the air filters in place.
Allow the air filters to dry throughly before re-installing them on
the unit.