EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ C9 Series

GE LOGIQ C9 Series User Manual

GE LOGIQ C9 Series
247 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
Environmental Dangers
LOGIQ C9 Series – Basic Service Manual 2-23
5482298-100 English Rev. 5
Patient Environment IEC60601-1 and ANSI AAMI ES60601-1
Figure 2-5. Patient environment
Sub Clause 3.79 and figure A.9 (IEC60601-1:2005 and ANSI AAMI
ES60601-1:2005)
Such an area is an environment in which medical diagnosis, monitoring or treatment is
carried out. It is very difficult to attach unique dimensions to the PATIENT
ENVIROMENT.
In practice a distance of 2,5 m (8.2 ft.) above the floor on which the medical personnel
stand and a horizontal distance of 1,5 m (4.9 ft.) have justified themselves as
indicative of the dimensions of the Patient Environment.
The patient environment/vicinity will be depicted as a dashed line in this procedure.
See example below.
1. Patient environment

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ C9 Series and is the answer not in the manual?

GE LOGIQ C9 Series Specifications

General IconGeneral
BrandGE
ModelLOGIQ C9 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals