Tools required
LOGIQ C9 Series – Basic Service Manual 10-11
5482298-100 English Rev. 5
Special tools, supplies and equipment used for maintenance
Table 10-4: Overview of tool requirements for periodic maintenance
Tool / kit Part Number Comments
Digital Volt Meter (DVM)
Anti Static Kit 46–194427P231
46–194427P279
46–194427P369
46–194427P373
46–194427P370
Kit includes anti–static mat, wrist strap and
cables for 200 to 240V system
3M #2204 Large adjustable wrist strap
3M #2214 Small adjustable wrist strap
3M #3051 conductive ground cord
Anti Static Vacuum Cleaner 46–194427P278
46–194427P279
120V
230V
Safety Analyzer The safety Analyzer tool should be calibrated
and compliant with AAMI/ESI 1993 or IEC
60601 or AS/NZS 3551.
QIQ Phantom E8370RB RMI Grayscale Target Model 403GS
NOTE! The use of a Phantom is not
required during Preventive Maintenance.
Customer may use it as part of their Quality
Assurance Program tests.
B/W Printer Cleaning Sheet See printer user manual for requirements
Color Printer Cleaning Sheet See printer user manual for requirements
Disposable Gloves