DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
3 - 20 Section 3-4 - Completing the setup
To disconnect a probe:
1.) Select a different probe or freeze the image before removing a probe in order to avoid disconnecting
a live probe.
2.) Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector.
3.) Remove the connector from the port.
4.) Ensure that the probe head is clean before placing the probe in its storage case, see: 10-5-9
"Cleaning" on page 10-12 for cleaning instructions.