DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 244 Section 8-10 - Casters and Brakes replacement
8-10-6 Pedal Mechanism Cam FRU replacement
Material Safety Data Sheet link for LOCTITE 290:
• Copy URL into browser for Healthcare SDS System:
-
http://gensuite.ge.com/geh/ehs/msds/msds.cfm?sno=W.137-337
Table 8-267 Manpower / Time and Tools
Manpower /
Time
Tools / PPE
One person /
30 minutes
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
8-2-6 "PPE Required During Service" on page 8-6.
(Tools):
• Torx driver - T30. Optional Torx L-wrench or Torx socket.
• Open end Wrench, 10 mm.
• Wrench, Box, 10 mm. Optional deep socket.
• Adjustable wrench, 152 mm (6 inches) long (if a longer adjustable
wrench is used, it will contact the floor when trying to turn it.
• Two M5 X 0.8 X 16 mm screws (P/N 46-312358P52) from Box 1 of
the LOGIQ E10 Hardware Kit (P/N 5310019-2).
• LOCTITE 290 (LOCTITE P/N 29031). See:
"Material Safety Data Sheet link for LOCTITE 290:" on page 8-244 to access data
sheet.
LOCTITE 290 can be obtained from Grainger.
• Grainger item number 5E217.
NOTE: An adjustable wrench is deeper than an open end wrench and will secure
the Cam Plate better in this procedure.