DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
1 - 24 Section 1-4 - Safety considerations
1-4-4 Electrical safety
Safe practices
Follow these guidelines to minimize shock hazards whenever you are using the LOGIQ E10:
• To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
• The LOGIQ E10 is equipped with a three-conductor AC power cable. This must be plugged into an
approved electrical outlet with safety ground. A separate power outlet with a 20 amp circuit breaker
for 120 VAC for 120V area, 7.5 amp circuit breaker for 220-240 VAC for 220/240V area or 15 amp
circuit breaker for 100 VAC for Japan.
• The power outlet used for this equipment should not be shared with other types of equipment.
• Both the LOGIQ E10 power cable and the power connector must meet international electrical
standards.
Probes
Follow these guidelines before connecting a probe to the LOGIQ E10:
• Inspect the probe prior to each use for damage or degradation to the:
- housing
- cable strain relief
-lens
- seal
- connectors
- locking mechanism
• Do not use a damaged or defective probe.
• Never immerse the probe connector or adapter into any liquid.
• The LOGIQ E10 has more than one type of probe port. Use the appropriate probe port designed for
the probe you are connecting.
Peripherals
• Refer to the “Peripherals used in the patient environment” in Chapter 2 of the User Manual for
peripheral isolation information.
WARNINGWARNING
Connecting the LOGIQ E10 to the wrong voltage level will most likely destroy it.