8-7-24 Storage Tray / Transvaginal Probe Holder replacement
The Storage Tray and Transvaginal Probe Holder rests inside the Options Holder base and are
interchangeable. They can also replace the Gel Warmer. See: 8-7-23 "Gel Warmer replacement" on
page 8-173 and 8-7-25 "Replacing the Gel Warmer with a Storage Tray or TV/TR Probe Holder" on page
8-179.
Follow this procedure for replacement, or replacing a tray with a Probe Holder or a Probe Holder with a
tray.
Storage Tray / Transvaginal Probe Holder removal
Table 8-209 Storage Tray / Transvaginal Probe Holder removal
Steps Corresponding Graphic
1. Locate the two prongs below the tray or
Probe Holder.
2. Squeeze the two prongs together, then lift
the tray or Probe Holder off the
LOGIQ E10.