DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 328 Section 8-11 - Front End Acquisition / Card Cage parts replacement
Rear I/O replacement
Table 8-338 Rear I/O replacement
Steps Corresponding Graphic
1.
The Rear I/O mounts directly to the Option
Box.
Located the two screws that secure the
Rear I/O to the Option Box, above the
Batteries.
Remove the screws using a #2 Phillips
screwdriver.
2.
Slide the Rear I/O straight off the Option
Box and disconnect the Network and
Power Cables.
3.
Slide the replacement Rear I/O straight onto the Option Box and re-connect the Network and
Power Cables.
4.
Re-install all Covers removed.