Connectivity setup
LOGIQ V2/LOGIQ V1 – Basic Service Manual 3-37
5610739-100 English Rev.10
Setup the LOGIQ V2/LOGIQ V1 for DICOM Communications
(continued)
Table 3-12: Utility ->Connectivity ->Device screen
1Select MyComputer.
2 Assign an AE title to the
LOGIQ V2/LOGIQ V1. (AE stands for Application Entity.
DICOM services use this to identify the
LOGIQ V2/LOGIQ V1.) AE title is
case-sensitive. This title may contain the Computer Name from the TCP/IP page,
if desired.
Note: It is NOT recommended to use the factory default. This is not prohibited, but
more than one system with the same AE title can cause confusion.
3 Edit Port Number if needed. 104 is typical. Save your changes and reboot the
system.