Functional checks
LOGIQ V2/LOGIQ V1 – Basic Service Manual 4-39
5610739-100 English Rev.10
Back End Processor checks
• If all the previous tests have been passed successfully, the
Back End Processor is most likely OK.
If the system seems to be operating erratically, please refer
to ‘Diagnostics/Troubleshooting’ on page 8-1.
Operator Panel Test
• The Operator Panel is tested when the LOGIQ V2/LOGIQ
V1 is powered up as part of the start-up scripts, run at every
start-up.
For more info, please refer to ‘Diagnostics/Troubleshooting’
on page 8-1.
Peripheral checks
Printer checks
The internal printer is controlled from the Print and Store keys
on the LOGIQ V2/LOGIQ V1’s Operator Panel.
The factory default is:
• Print for the standard printer
Store for the screen capture to clipboard
Turn OFF Power to LOGIQ V2/LOGIQ V1
See ‘Power off’ on page 4-9 for more information.