Completing the setup
LOGIQ V2/LOGIQ V1 – Basic Service Manual 3-15
5610739-100 English Rev.10
Electrical specifications
Verification of the system’s voltage setting
Verify that the mains voltage specified for the LOGIQ V2/LOGIQ
V1 is available on-site.
Refer to the latest revision of the User Manual for a full list of
product labels prior to serving the system.
Electrical specifications for LOGIQ V2/LOGIQ V1
In the table below, the electrical specifications for LOGIQ V2/
LOGIQ V1 includes monitor and on board peripherals.
Connecting a LOGIQ V2/LOGIQ V1 to the wrong voltage level
will most likely destroy it.
Table 3-3: Electrical specifications for LOGIQ V2/LOGIQ V1
Part
Number
Description Voltage Tolerances
Power
consumntion
Frequency
5661244 LOGIQ V2 Generic
version
100-240V ±10% Max.200VA 50/60 Hz
5661242 LOGIQ V1 Generic
version