EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V2

GE LOGIQ V2 Service Manual

GE LOGIQ V2
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #295 background imageLoading...
Page #295 background image
System maintenance
LOGIQ V2/LOGIQ V1 – Basic Service Manual 10-19
5610739-100 English Rev.10
Battery Performance Maintenance
Battery replacement every three years is recommended.
It is recommended to do battery performance maintenance one
time per year.
Please follow the flow chart below to carry out battery
performance maintenance.
Figure 10-1. Flow chart of Battery Performance Maintenance
NOTE: Disconnect all probes when discharge battery.
NOTE: Discharge the battery to let the system automatically shut down.
Battery discharge
Battery charge > 3 hour
Yes
No
Battery power > 90%?
Battery charge > 3 hour
Battery power > 90%?
No
Battery is Ok
Yes
Yes
Discharge Replace battery
No
Yes
Battery is Ok
Replace battery
>45 minutes?
Replace battery
>45 minutes?
No

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals