System maintenance
LOGIQ V2/LOGIQ V1 – Basic Service Manual 10-13
5610739-100 English Rev.10
Functional checks
NOTE: See also Chapter 4
The functional checks take about 60 minutes to perform. Refer
to the Ultrasound system user documentation whenever
necessary.
System checks
Table 10-6: System functional checks
Step Item Description
1. B-Mode Verify basic B-Mode operation. Check the basic Ultrasound system
controls that affect this mode of operation.
2. CF-Mode Verify basic CF-Mode (Color Flow Mode) operation. Check the basic
Ultrasound system controls that affect this mode of operation.
3. Doppler Modes Verify basic Doppler operation (PW and CW if available). Check the basic
Ultrasound system controls that affect this mode of operation.
4. M-Mode Verify basic M-Mode operation. Check the basic Ultrasound system
controls that affect this mode of operation.
5. Probe Elements Perform an Element Test on each probe to verify that all the probe
elements and system channels are functional.
6. Applicable
Software Options
Verify the basic operation of all optional modes such as Contrast. Check
the basic Ultrasound system controls that affect each options operation.
7. Xmit/Recv
Elements
Use the Visual Channel Utility on the loop connect to verify that all system
xmit/recv channels are functional.
8. Operator Panel
test
Perform the Operator Panel Test Procedure.
9. Keyboard Do the interactive keyboard test.
10. LCD Verify basic LCD display functions. Refer to Chapter 3 of the User Manual.
11. Software Menu
check
Verify Software Menu display functions. Refer to Chapter 3 of the User
Manual.
12. Peripherals See: ‘Peripheral checks’ on page 4-39.
13. Measurements In measurement mode, make distance measurement, get result in result
window. Verify the distance by graduate rule. Distance Accuracy should
be within ±5%. (Name result from result window Result A, result from
graduate rule Result B; Distance Accuracy = (Result B-Result A)/Result A)

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals