General Procedures and Functional Checks
4-38 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Cineloop Check
Introduction
A cineloop is a sequence of images recorded over a certain time
frame. When using ECG the time frame can be adjusted to
cover one or more heart cycles. When frozen, the System
automatically displays the cineloop boundary markers on either
side of the last detected heart cycle.
Preparation
• Connect one of the probes to the scanner.
• See ‘Probe/Connectors Check’ on page 4-37 for info about
connecting the probes
For available probes, see ‘Probe and Accessories’ on
page 9-9:
• Turn ON the scanner. The 2D Mode is displayed (default
mode).
Adjust the Cineloop controls
• Press Freeze.
The left and right markers are displayed on either side of the
last detected heart cycle on the ECG trace.
• Press Freeze.
The selected heart beat is played back.
• Press Freeze to freeze the cineloop.
Use the trackball to scroll through the acquisition and find
the sequence of interest.
• Adjust Cycle select to move from heart beat to heart beat
and select the heart cycle of interest.
Adjust Num cycles to increase or decrease the number of
heart beats to be played back.
Adjust Left marker and Right marker to trim or expand the
cineloop boundaries.

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals