EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V2

GE LOGIQ V2 Service Manual

GE LOGIQ V2
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
General Ultrasound system requirements
LOGIQ V2/LOGIQ V1 – Basic Service Manual 2-3
5610739-100 English Rev.10
General Ultrasound system
requirements
Contents in this section
• ‘Ultrasound system environmental requirements’ on
page 2-3
• ‘Electrical requirements’ on page 2-6
• ‘EMI limitations’ on page 2-9
• ‘Probes environmental requirements’ on page 2-11
• ‘Time and manpower requirements’ on page 2-11
Ultrasound system environmental requirements
If the Ultrasound system is very cold or hot
When unpacking the Ultrasound system, allow the temperature
of the Ultrasound system to stabilize before powering up. The
following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
CAUTION
If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Table 2-1: System Acclimation Time Chart
Degree C 50 45 40 35 30 25 20 15 10 5 0 -5
Degree F 122 113 104 95 86 77 68 59 50 41 32 23
hours
4
2
0
000000246

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals