Maintenance
VisualInspections
Performtheappropriatevisualinspectionsfortherepairprocedureperformed.
1.Inspectthepatientcable,leadwires,andelectrodes.
See“VisualInspection”onpage137formoreinformation.
2.Discusselectrodeplacement,skinpreparation,andpatient-relatedrequirements
withtheECGtechnician.Verifythecustomerisfollowingtherecommended
procedures.
RefertoChapter4,PreparingthePatient,intheMAC™2000ECGAnalysisSystem
Operator’sManualformoreinformation.
3.VerifythekeypadandLCDdisplaypassesvisualinspection.
See“VisualInspection”onpage137formoreinformation.
4.VerifytheACpowercordpassesvisualinspection.
See“VisualInspection”onpage137formoreinformation.
5.Verifythebatterypackpassesvisualinspection.
See“VisualInspection”onpage137formoreinformation.
6.VerifytheSDcardpassesvisualinspection.
See“VisualInspection”onpage137formoreinformation.
7.Verifythatallharnessesandinternalwiringpassvisualinspection.
See“VisualInspection”onpage137formoreinformation.
8.Verifythatallfastenershavebeenreplacedandsecured.
See“VisualInspection”onpage137formoreinformation.
FunctionalCheckoutProcedures
Performtheappropriatefunctionalcheckoutprocedurefortherepairprocedure
performed.
OperationalChecks
1.Performthepower-upself-testandverifythesystempasses.
Ifasoftwareupdatehasbeenperformed,verifythenewversionofsoftwareis
identiedontheMainMenuscreen.
See“PowerUpSelfTest”onpage136formoreinformation.
2.Recordarhythmstripandverifyitrecordedsuccessfully.
RefertoChapter7,RecordingaRestingECG,intheMAC™2000ECGAnalysis
SystemOperator’sManualformoreinformation.
3.RecordanECGandverifyitrecordedsuccessfully.
RefertoChapter7,RecordingaRestingECG,intheMAC™2000ECGAnalysis
SystemOperator’sManualformoreinformation.
4.StoreanECGandverifyitstoredsuccessfully.
RefertoChapter11,ManagingInternalStorage,intheMAC™2000ECGAnalysis
SystemOperator’sManualformoreinformation.
5.TransmitasimulatedECGandverifythedatawassuccessfullyreceived.
RefertoChapter11,ManagingInternalStorage,intheMAC™2000ECGAnalysis
SystemOperator’sManualformoreinformation.
6.Connectexerciseequipmenttothesystemandverifytheequipment
communicatessuccessfullywiththesystem.
2053535-003C
MAC™2000ECGAnalysisSystem
131