ServiceSetup
Afterthedevice’sashmemoryisformatted,youneedtorecongurethedevice,
RefertotheMAC™2000ECGAnalysisSystemOperator’sManualforinformationon
conguringthedevice.
Ifformattingthedevice’smemoryashdoesnotresolvethememoryordata
corruptionissues,youneedtoreplacethemainPWA.Forinstructionsonreplacingthe
PWA,see“ReplacingthePWA”onpage124.
OpeningaCommandPrompt
Usethefollowingprocedurestoopenacommandpromptfromwhichyoucanissue
variousnetworkdiagnosticcommands.
1.GototheServiceSetupmenu,asdescribedin“AccessingServiceSetup”on
page43.
2.SelecttheOpenCommandPromptbutton.
Thefollowingcommandpromptwindowopens:
3.Issueanyofthefollowingcommandsasnecessary.
Command
Description
ping<ipaddress>
ThesystemqueriesandreceivesrepliesfortheIPaddress
identiedinthecommand.
ipconfig
ReturnsinformationrelatedtotheTCP/IPcongurationof
thesystem.
netstat
Returnsnetworkstatisticsforthesystem.
netstat-e
Returnsethernetstatisticsforthenetwork.
netstat-n
Returnssystemnetworkaddressesandportnumbersin
numericalform.
netstat-p
ReturnsstatisticsspeciedfortheIP4protocol.
netstat—r
Returnstheroutingtable.
4.Whenyouaredone,typeexitandpressEntertoclosethecommandprompt
windowandreturntotheServiceMenu.
70
MAC™2000ECGAnalysisSystem
2053535-003C

Table of Contents

Other manuals for GE MAC 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 2000 and is the answer not in the manual?

GE MAC 2000 Specifications

General IconGeneral
BrandGE
ModelMAC 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals