ElectromagneticCompatibility(EMC)
EMCImmunityTest
ImmunityTest
ComplianceElectromagnetic
Environment–Guidance
Electrostaticdischarge(ESD)
•IEC60601-1-2:2007
•IEC61000-4-2:2001
±2/4/6kVindirect
±2/4/6kVdirect
±2/4/8kVair
Exception:
±2/4kVairforPaperQueue
sensors
Floorsshouldbewood,
concrete,orceramictile.
Iftheoorsarecovered
withsyntheticmaterial,the
relativehumidityshouldbe
atleast30%.
ElectricalFast
Transient/burst(EFT)
±2kVforpowersupplylines
±1kVforinput/outputlines
Mainspowershouldbethat
ofatypicalcommercialor
hospitalenvironment.
FastTransientSurge(FTS)
±1kVdifferentialmode
±2kVcommonmode
Mainspowershouldbethat
ofatypicalcommercialor
hospitalenvironment.
Voltagedips,short
interruptions,andvoltage
variationsonpowersupply
inputlines
<5%Ut(>95%dipinUt)for
0.5cycles
<40%Ut(>60%dipinUt)for
5cycles
<70%Ut(>30%dipinUt)for
25cycles
<5%Ut(>95%dipinUt)for
5s
Mainspowershouldbethat
ofatypicalcommercialor
hospitalenvironment.
Iftheuserrequirescontinued
systemoperationduring
powermainsinterruptions,
itisrecommendedthat
thesystemispowered
fromanapplicablyrated
uninterruptiblepowersupply
orabattery.
Powerfrequency(50/60Hz)
magneticeld
3A/m
Powerfrequencymagnetic
eldsshouldbeatlevels
characteristicofatypical
locationinatypical
commercialorhospital
environment.
NOTEU
t
istheACmainsvoltagepriortoapplicationofthetestlevel.
GuidanceandManufacturer’s
Declaration—ElectromagneticImmunity
Thesystemdescribedinthismanualisintendedforuseinthefollowingspecied
electromagneticenvironment.Itistheresponsibilityofthecustomerorusertoensure
thatthissystemisusedinsuchanenvironment.
2053535-003C
MAC™2000ECGAnalysisSystem
167