TechnicalSpecications
Processing(cont'd.)
Item
Specications
ECGtransmission
RestingECGscanbetransferredviatheserialinterface,modem,
LAN,aswellasSD-cardfromtheMAC2000systemtootherECG
systems(CardioSoft/CardioSys,MUSE)forfurtherprocessingand
archiving.
Inaddition,RestingECGcanbereceivedviaserialline,modem,
andSDCard.
PatientData
Patientinformationentry:PatientID,SecondaryPatientID,
Height,Weight,Gender,Race,PacemakerPatient,SystolicBP,
DiastolicBP,Location#,Room,OrderNumber,PhoneNumber,
Medication,OrderingPhysician,ReferringPhysician,Attending
Physician,Technician,Testindication,fouruser-denableelds.
Electrodeconnections
RA,RL,LA,LL,V1–V6,Nax,Nst
Detectionofpacemaker
pulse
Lowerscale:0.2msduration,0.5mVamplitude
Higherscale:2.0msduration,250mVamplitude
ArchivingECGdataisprintedout
Internalmemoryispresentonthissystem.
ExternalStorageSDCardavailableforstorage
Input/OutputInterface
RS232portforconnectivitytoMUSEsystemsv7.1.1andv8.0.1
andCASE/CardioSoftsystemsV6.51,V6.61,andCardioSoftV6.71,
andremotecontrolsignalfromergometers
Ethernetinterface
RJ45connector
Channels
6,12
QRSdetectionAvailable
Musclelter
20Hz,40Hz,100Hz,150Hz
Automaticbaseline
correction
User-selectable
OperatingModes,Features,andOptions
OperatingModesandFeatures
Item
Specications
RestingECGModeRecordsandprints12-leadrestingECGswith10secondsduration
asastandardfeature.
ArrhythmiaModeContinuouslymonitorsECGandprintsreportwhenarrhythmia
eventsoftheuser-selectedclassoccur
Exercise/StressTest
Mode
Exercisemodeforexercisestresstesting
RRAnalysisModeRRAnalysisforRRintervalanalysis
HookupAdvisorProvidesvisualindicationofsignalquality
2053535-003C
MAC™2000ECGAnalysisSystem
175