TechnicalSpecications
ECGDataAcquisition(cont'd.)
Item
Specication
Commonmoderejection
>135dBwith50/60Hzlteron
Inputimpedance
>10@10Hz
Patientleakagecurrent<10µAnormalcondition(NC)
<50µAsinglefaultcondition(SFC)
RemotecontrolAdvancingtothenextstagecanbeinitiatedfromthebicycle
ergometer.
Processing
Item
Specications
Processor
Application:32bitARM9processor
PeripheralManagement:32bitARMCortex-M3processor
SignalChainProcessing:400MHzDSP
ECGInterpretation
Marquette12SLECGAnalysisProgramforadultsandpediatrics
HeartAlgorithm(Stress)
Computerized
measurements
12-leadanalysis
ECGanalysisfrequency
500or1000samples/second/channel
Digitalsamplingrate
16000samples/second/channelfornormaldataacquisition
Pacesamplerate
75Ksamples/second/channel
ECGon-screenpreview
On-screenpreviewofacquired10-secondECGwaveform
AcquisitionmodePre-acquisitionorpostacquisition,provides10secondsof
instantaneousECGacquisition
Resolution
4.88µV/LSB
Frequencyrange
0.04to150Hz
Lowcutofffrequency0.04Hz(-3dBlimits)
HighcutofffrequencyCongurableat20Hz,40Hz,100Hz,150Hz
Commonmoderejection
>135dB(with50/60Hzlteron)
Inputimpedance>10@10Hz(debrillatorprotected)
Patientleakage
<10µANC
<50µASFC
Electrodesmonitoredfor
disconnection
EveryelectrodeexceptRLandRA
Heartratemeter
30to300BPM
Operatingsystem
MicrosoftWindowsCE6.0
Startuptime
Lessthan30seconds.
174
MAC™2000ECGAnalysisSystem
2053535-003C

Table of Contents

Other manuals for GE MAC 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 2000 and is the answer not in the manual?

GE MAC 2000 Specifications

General IconGeneral
BrandGE
ModelMAC 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals