7. Specifications
Page 172 DMS 2 Operating Manual
7. Specifications
7.1 DMS 2 Specifications
Operating Principle: Ultrasonic pulse/echo
Measurement Modes: Dual-element, Dual (D-MULTI) & Single-element/Multiple-echo, SIP, B-scan
Probe Zero Adjustment: Automatic; Push button, keyed to instrument probe zero block
V-Path Correction: Automatic, microprocessor controlled
Linearity: Dual: ±.002 inch (±.05 mm) when calibration point < 1 inch (25 mm);
±.003 inch (.075 mm) when calibration point > 1 inch (25 mm)
Single: ±.001 inch (±.025 mm)
Measuring Range: .008 to 25.00 inch (.2 to 635 mm) depending on probe and material
Digital Displayed Resolution: .001 or .01 inches
.01 or .1 mm
Update Rate: Selectable 4, 8, 12, or 16 Hz: 32 Hz in Min Cap and B-scan Modes
Material Velocity Range: 39,400 to 393,662 in./sec.; 1000 to 9999 m/sec.
Display Type: High Resolution: ¼ VGA, 240 X 320 Pixels, Graphic Waveform LCD, 2.8" X 3.75"
(71 X 95 mm) viewing area, backlight, adjustable contrast
Thickness Value Display: 5 digit, .5" (10.7 mm) high; .75” digit height; In Big Digits Mode