7. Specifications
Page 178 DMS 2 Operating Manual
7.2.3 Recommended probes for the D-Multi Mode
Model Designation Nominal Contact Temperature Measuring Range
Frequency Diameter Range in Steel
HT400A/400 High Temperature 5 MHz .500” <1000°F .100”-2.000”
(12.7mm) (<538°C) (2.5-50mm)
KBA560-D General Purpose 5 MHz .625” <450°F .100”-2.000”
(15.9mm) (<230°C) (2.5-50mm)
DA301 General Purpose 5 MHz .475” <140°F .070”-1.800”
(12.1mm) (<60°C) (1.8-46mm)
DA312 Thin Materials 10 MHz .300” <130°F .030”-.900”
(7.6mm) (<54
°C) (.8-23mm)
DP-104 High Penetration 1 MHz 1.25” <130°F .100”-.2000”
(31.8mm) (<54°C) (2.5-50mm)
DM-401-GP General Purpose 5 MHz .700” <300°F .060”-8.00”
(17.8mm) (<150°C) (1.5-200mm)
Note: Other probes listed in the D-Multi Mode probe parameter in the TG Setup Menu are compatible
over a limited thickness range.
NOTE:
Probe specifications are subject to change without notice.