7. Specifications
Page 176 DMS 2 Operating Manual
Dual Element Probes (continued)
Nominal Contact Measurement Temp.
Model Probe Description Frequency Diameter Range (Steel) Range Cable
DP-104 High Penetration 1 MHz 1.25" .200" min <130°F KBA-532
31.8 mm 5.0 mm min <54°C
DA301 General Purpose 5 MHz .475" .050" to 8.0" <130°F KBA-533
12.1 mm 1.25 to 200 mm <54°C
DA303 High Penetration 2 MHz .635" .200" min <130°F KBA-533
16.2 mm 5.0 mm min <54°C
DA312 Thin Materials 10 MHz .30" .025" to 2.0" <130°F KBA-532
7.6 mm .64 to 50 mm <54°C
KBA550BTH Boiler Tube 5 MHz .375" .060" to 2.0" <130°F C-BTH
9.5 mm 1.5 to 50 mm <54°C
KBA550FH Fingertip 5 MHz .375" .060" to 2.0" <130°F Potted
9.5 mm 1.5 to 50 mm <54
°C
KBA525 5 mm Footprint 10 MHz .200” .025” to 1.0” <130
o
F Potted
5.0 mm .6 to 25 mm <54
o
C
DA312B16 3 mm Footprint 10 MHz .120” .025 to .500” <130
o
F Potted
3 mm .6 to 13 mm <54
o
C
TC-560 TopCOAT/Auto-V 5 MHz .675” 0.002 to 0.080”* <130°F KBA-532 or
15.9 mm (.05 to 2.00 mm) <5C KBA-531A
* Measurement range for coating. 0.060 to 8.00”**
** Measurement range for material. (1.5 to 200 mm)
NOTE:
Probe specifications are subject to change without notice.