IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
6 - 10 Section 6-5 - Touch Screen Calibration
Section 6-5
Touch Screen Calibration
The Touch Screen Calibration is found on the Operator Panel Test Dialog.
Follow these steps to open the Operator Panel Test Dialog:
1.) Select Config (F2).
2.) Log on as ADM.
3.) Select System > Test.
4.) Select Operator Panel Test. This will open the Operator Panel Test Dialog (Figure 6-9).
The Calibrate button is used to start to calibrate the position of the touch sensitive area on the Touch
screen relative to the graphic artwork.
Follow these steps to calibrate the Touch screen:
1.) Select Calibrate. The Touch screen goes blank and display a cross in the screen’s upper left
corner.
2.) Point your finger on the centre of the cross and press slightly. This calibrates this point, and the
cross moves to the screen’s upper right corner.
3.) Point your finger on the centre of the cross and press slightly. This calibrates this point, and the
cross moves to the lower right corner.
4.) Point your finger on the centre of the cross and press slightly. This calibrates this point, and the
cross moves to the lower left corner.
5.) Point your finger on the centre of the cross and press slightly. This calibrates this last point.
6.) Select Done. The calibration is complete.
Figure 6-9 The Operator Panel Test Dialog

Table of Contents

Other manuals for GE Vivid E7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E7 and is the answer not in the manual?

GE Vivid E7 Specifications

General IconGeneral
BrandGE
ModelVivid E7
CategoryMedical Equipment
LanguageEnglish

Related product manuals