IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
7 - 2 Section 7-2 - Service Safety Considerations
Section 7-2
Service Safety Considerations
Section 7-3
FAQ - Frequently Asked Questions
7-3-1 High System Temperature Error
Question:
What is the highest temperature when VIVID E9 / VIVID E7 starts giving the “High System
Temperature” Error?
Answer:
• All Card Rack (PCBs, or Front End) sensors:
- The limit for the first warning is 78 degrees centigrade (78 ºC).
- The system starts the shutdown sequence at 80 degrees centigrade (80 ºC)
• DRX_TOP sensors:
- The limit for the first warning is 87 degrees centigrade (87 ºC).
- The system starts the shutdown sequence at 90 degrees centigrade (90 ºC).
• Clean air filter:
- The reminder “Clean air filter” is displayed at start-up of the VIVID E9 / VIVID E7, and is
repeated at specific pre-defined intervals.
7-3-2 Reset the BEP from a Hang
Question:
It seems that the BEP is “hanging” - it’s not responding at all. How do I Reset it?
Answer:
1.) Try this method first:
Press the ON/OFF switch on the Operator Panel for more than six seconds. This should cause the
VIVID E9 / VIVID E7 to perform a “Forced Restart”.
2.) If the instruction in step 1 didn’t help, as the last solution, after waiting several minutes, switch off
the power on the rear of the system.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
IF THE COVERS ARE REMOVED FROM AN OPERATING VIVID E9 / VIVID E7, SOME
METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD
IF TOUCHED, EVEN WHILE IN SHUT DOWN MODE.
USE ALL PERSONAL PROTECTION EQUIPMENT (PPE) SUCH AS GLOVES, SAFETY
SHOES, SAFETY GLASSES, AND KNEELING PAD, TO REDUCE THE RISK OF INJURY.