GE
D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 8 - Replacement procedures 8 - 3
8-2-2 Warnings (cont’d)
Spare Part order for Vivid E9 VET:
The same spare parts used in the current VIVID E9 consoles can be used in the
veterinary consoles.
When ordering any of the following, the Vet label(s) MUST be ordered as they need to
be applied to the new part:
- Side Covers - require label P/N: 5447716
- Back Cover - require label P/N: GB200182
- Probes - require label P/N: 5454608
When performing replacement procedures within a veterinary environment, return any used/
unused spare parts with the purple scrap marking label (regardless of the actual condition of
the part) and add a description on the label stating that the items were removed from a
veterinary environment.