IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 8 - Replacement procedures 8 - 23
8-4-12 Verifications after the software has been re-loaded
NOTE: DO NOT use this procedure if all partitions on the HDD have been erased during the software
load, but refer to: 8-4-11 "Setup after Software loading" on page 8-22.
NOTE: Use this procedure if only the software on C:\ has been updated or reloaded.
1.) With your recordings from before the software loading available, continue with the setup instructions
starting in: Section 3-6 "Configuration" on page 3-24. Correct the settings if needed.
2.) Continue with: Section 3-8 "Connectivity setup" on page 3-38. Based on your recordings, correct
the settings if needed.
3.) Continue with: Section 3-9 "Options Setup" on page 3-52. Based on your recordings, correct the
settings as needed.
4.) After a software load, you should always calibrate the Front End, as described in: Section 6-6 "DC
Offset Calibration (Front End Alignment)" on page 6-11.

Table of Contents

Other manuals for GE Vivid E7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E7 and is the answer not in the manual?

GE Vivid E7 Specifications

General IconGeneral
BrandGE
ModelVivid E7
CategoryMedical Equipment
LanguageEnglish

Related product manuals