Maintenance Procedures
032-0265-EN Rev 1 5-3
Cleaning EO/CH (Panoramic and Pephalometric)
Carousel
4. To clean the EO/CH carousel, you
will follow similar steps to the
intraoral carousel. Remove the
carousel
imaging plate holder by
removing the three screws of the
plastic retention piece on both
sides of the carousel. The imaging
plate holder will easily come free.
Spray the carousel with a general
househol
d nonabr
asive cleaning
solution. Do not use strong alkaline
or ammonia based cleaners. Wipe
clean and allow to air dry. If the
intraoral imaging plate holder is
soiled, remove it by removing both
vertical plastic retention pieces by
removing the three screws on each
side and following the same
instructions as for the intraoral
image plate holder cleaning.
Clean with the general household
non-
abras
ive spray or soap and
water. Allow to dry completely
before re-assembling. Re-assemble by placing one side of the imaging plate holder
under the plastic retention device and tighten screws. Repeat step on the other
side of the imaging plate holder, replace plastic retention device and tighten the
three screws.
Operator Maintenance
Scanner Maintenance
The only part of the DenOptix QST scanner that may require operator maintenance
is the fuse. If you experience any other service-related problem, contact your local
Gendex dealer. Follow these directions to change the fuse.