Guidance and Manufacturer’s Declaration - Electromagnetic Compatibility
Essential Performance: for the purpose of EMC compliance, no device functions are considered
essential to the safety of the patient.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Gendex QST is intended for use in the electromagnetic environment specified below. The
customer or the user of the Gendex QST should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The Gendex QST uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment
RF emissions
CISPR 11
Class A The Gendex QST is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies