105
Troubleshooting
LED indicators (wireless modules)
The wireless module’s LEDs are used to give you indications of the present operation status.
The following table describes the operation status and remedies against issues.
Power status
AC LED BATT LED Status
Lighting up. O Running on externally supplied power.
O Lighting up. Running on battery.
O Repeatedly emits 3 ashes every 3 s. Battery is low.
While the battery charging is in progress, the CHARGE LED remains on.
Measurement status
MEASURE LED Status
O Measurement stops.
Lighting up. Measurement in progress.
Emits 5 ashes for 1 s. Preparing.
Wireless communications status
CONNECT LED Status Remedy
O No communications in progress. –
Lighting up. Communicating with LR8450-01. –
Flashes in 2 s interval. Communications disrupted.
Reconnecting.
Please wait.
Flashes in 1 s interval. Connecting with auto-connect. Please wait.
Repeatedly emits 3 flashes
every 3 s.
Radio-wave intensity weakens. Place module in another place.
Emits 5 ashes for 1 s. Processing. Please wait. Have it repaired if not
remedied.
Error status
ERR LED Status Remedy
Rapidly blinking. Z3230 not attached. Attach Z3230 Wireless LAN Adapter.
Emits blinks in other patterns. Internal error occurs. Have it repaired.
Maintenance and Service