Installation SmartRadar FlexLine
Installation Guide Part No.: 4417.760_Rev06
40
4.11.4.3.3 MPT - MRT (Board ID = 4)
4.11.4.4 Stand-alone SmartView Display
The labelling on the wiring terminal in the SmartView is identical
to the labelling in the SmartRadar.
Label Connections diagram Parameters
38. S- [Ex ia] IIB/IIC
U
o
= 8.61 V
I
o
= 58 mA
P
o
= 0.13 W
C
o
= 1.0 μF
L
o
= 10.5 mH
39. com
40. R1
41. R2
42. R3
43. R4
44. R5
45. R6
Cable specifications
Type 4-wire, shielded
Maximum length 50 meter
Label Parameters
26. SGND [Ex ia] IIB
U
o
= 14.2 V
I
o
= 522 mA
P
o
= 1.7W
C
o
= 2.18 μF
L
o
= 0.25 mH
27. Vsafe_+
28. Safe_A
29. Safe_B
R
c
R1
R2
R3
R4
R5
R6