EasyManuals Logo
Home>Honeywell>Transmitter>XNX

Honeywell XNX User Manual

Honeywell XNX
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
XNX Universal Transmitter Quick Start Guide
48
3. IftheZeroCalibrationissuccessful,theXNXUniversalTransmitter
willdisplaytheZeroPassedscreen.
Span Calibration
NOTE:
If a Span Calibration is not required, select the
to skip the Span
Calibration and return to the Calibration menu.
4. WhentheZeroCalibrationiscompleteoritisskipped,theSpan
Concentrationscreenappearstoindicatetheconcentrationvalue
ofthegasusedforcalibration.
Figure 45. Span Gas Concentration screen
5. Select
tochoosetherstdigitandusethe switchestoincrement
ordecrementthevalues.Select
toacceptthenewvalueandmoveto
thenextdigit.Continueuntilall3digitshavebeenselected.
Sensor Reading
at Current Setting
s
Calibration Gas
Concentration
Figure 46. Span Calibration screen
6. Select
thenapplythetargetgas.Asthesensordetectsthegas
andtheconcentrationincreases,thesensorreadingvaluesinthe
displaywillchangetoreectthechangingconcentration.
7. Whentheconcentrationvaluesstabilize,thegasconcentration
asreadbytheinstalledsensorisstable.Atthistime,thegas
readingsaretakenbythesensor.TheSpanCalibrationprocess
alsodetermineswhetherthesensoriswithintheproperrangeto
accuratelydetectthetargetgas.
8. Whenthesensorhascompletedthecalibrationandthespan
algorithmshavedeterminedthatitiswithinrange,theSpanPassed
screenwillappear.
If the calibration is not successful, the Span Failed screen will display. Selecting
will return to the Span Concentration screen to begin the span calibration
again.
will exit Span Calibration and return to the Main Calibrate screen.
Arrow indicates
gas values too low
Selecting ‘ returns to
the Span Concentration
screen
Figure 47. Span Calibration Failed screen
Once the Zero and Span calibrations are completed successfully, the XNX will exit
the calibration procedure. Before returning to the Gas Calibration menu however,
the user will be prompted to exit with inhibit off, exit with inhibit on, or not exit.

Table of Contents

Other manuals for Honeywell XNX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XNX and is the answer not in the manual?

Honeywell XNX Specifications

General IconGeneral
BrandHoneywell
ModelXNX
CategoryTransmitter
LanguageEnglish

Related product manuals