XNX Universal Transmitter Quick Start Guide
47
8.1 Calibration
WARNING
Do not use the XNX Universal Transmitter in oxygen-enriched atmospheres.
Concentrations displayed will be adversely aected by oxygen depletion.
CAUTION
The calibration procedure should only be performed by qualied personnel.
NOTE:
The default calibration values for the “Calibration Required” diagnostic
vary based on sensor type. This value can be reprogrammed in
accordance with site requirements to ensure the highest level of safety.
Correct operation of each sensor/detector should be conrmed using
calibration with a certied test gas of known concentration before
commissioning. See Section 9 - Sensor Data for calibration gas
specications.
8.1.1 Calibration Procedure
NOTE:
Follow the specic procedure outlined in the Operating Manual for each
sensing device. The Zero Calibration procedure should be performed
prior to the Span Calibration procedure.
1. Ifusingcompressedgascylinder,attachthecalibrationgasow
housingontothebottomofthesensorandapplythegas.
2. Accessthecalibrationmode.TheGasCalibrationmenuisforboth
ZeroandSpanCalibration.
Figure 42. Gas Calibration menu
ZeroCalibration
Sensor Reading
Figure 43. Zero Calibration screen
Figure 44. Zero Calibration in
Progress
Select
✓
then apply the zero gas. As the sensor
detects the gas and the concentration increases, the
values displayed will reflect the changing
concentration. Selecting
✖
will return to the Gas
Calibration menu.