Functions and Features
4.7 List Function
The user can use list function to create a test program made of multiple steps.
Up to 10 programs (List1 to List10) can be created in the list mode for this series
power supplies. Each list program can be made of up to 100 steps. The user
need to set voltage or current, slope and dwell time for each step. The user also
can set repeat times, end state and so on for each list program.
In the channel combination mode Combine, the List function is
unavailable.
The descriptions of list function menu are listed in the table below.
List Mode List mode
Run List Turn the list mode on/off.
On Turn the list mode on.
Off Turn the list mode off.
Apply
Channel
Select channel as CH1 or CH2
Recall List
Select a list program to view.
List Setup View the setting parameters of the selected list
program.
View Step Select the desired step to view.
Edit List
Edit list program.
List Setup Set the list program.
Mode Set the operation mode for list program.
• CC: current mode;
• CV: voltage mode.
End State Set the running state after the list pro-
gram is running over.
• Normal: keep outputting the voltage
and current setting in main interface
when the dwell time is over.
• Last: keep outputting the last step
when the dwell time is over.
Repeat Set the number of loops to run the List
file, range: 1–65535. Set to 65535 to run
in an infinite loop.
Edit Step Edit parameters for each step.
Copyright © Itech Electronic Co., Ltd.
70