EasyManuals Logo
Home>Keithley>Multimeter>2110-100

Keithley 2110-100 User Manual

Keithley 2110-100
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
Model 2110
5½ Digit Multimeter Reference Manual Section 5: Remote co
mmands
2110-901-01 Rev. C/August 2013 5-39
[SENSe:]CURRent[:DC]:RESolution <n>|MIN|MAX
Use this command to select the resolution for DC current measurements (DCI).
When you set the resolution, the instrument automatically updates the integration time.
Parameters
<n> description
(3.00e-07 to 3.00e-04) * DCI
range
Select the resolution of the measurement.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]CURRent[:DC]:RESolution?
Query the resolution setting.
[SENSe:]CURRent[:DC]:RESolution? MIN
Query the minimum resolution setting.
[SENSe:]CURRent[:DC]:RESolution? MAX
Query the maximum resolution setting.
Details
Refer to Integration time and resolution for DC and resistance measurements (on page 3-41) more information
about resolution.
[SENSe:]CURRent:AC:RANGe:AUTO <b>
This command turns AC current autorange on or off.
Parameters
<b> description
0 or OFF
Turn autorange off.
1 or ON
Turn autorange on.
Query
command description
CURRent:AC:RANGe:AUTO?
Query autorange state.

Table of Contents

Other manuals for Keithley 2110-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110-100 and is the answer not in the manual?

Keithley 2110-100 Specifications

General IconGeneral
BrandKeithley
Model2110-100
CategoryMultimeter
LanguageEnglish

Related product manuals