EasyManuals Logo
Home>Keithley>Multimeter>2110-100

Keithley 2110-100 User Manual

Keithley 2110-100
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
Section
5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-48 2110-901-01 Rev. C/August 2013
[SENSe:]RESistance:RESolution <n>|MIN|MAX
Use this command to set a measurement resolution for 2-wire (Ω2) resistance measurements.
When you set the resolution, the instrument automatically updates the integration time.
Parameters
<n> description
(3.00e-07 to 3.00e-04) * Ω2 range
Select the resolution of the measurement.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]RESistance:RESolution?
Query the resolution setting.
[SENSe:]RESistance:RESolution? MIN
Query the minimum resolution setting.
[SENSe:]RESistance:RESolution? MAX
Query the maximum resolution setting.
Details
Refer to Aperture and resolution for frequency and period measurements (on page 3-45) for more information
about resolution.

Table of Contents