EasyManuals Logo
Home>Keithley>Multimeter>2110-100

Keithley 2110-100 User Manual

Keithley 2110-100
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #176 background imageLoading...
Page #176 background image
Model 2110
5½ Digit Multimeter Reference Manual Section 5:
Remote commands
2110-901-01 Rev. C/August 2013 5-69
TRIGger:DELay:AUTO <b>
Turn automatic trigger delay on or off.
The delay is determined by function, range, integration time, and AC filter setting. Specifying a delay time
automatically turns off the automatic trigger delay.
Parameters
<b> description
0 or OFF
Turn selected automatic trigger delay off.
1 or ON
Turn automatic trigger delay on.
Query
command description
TRIGger:DELay:AUTO?
Query the automatic trigger delay mode. Returns 0 (OFF) or 1 (ON).
TRIGger:SOURce <name>
Use this command to select a trigger source. The instrument accepts a software (BUS) trigger, an immediate
internal trigger, or a hardware trigger from the rear-panel EXT TRIG terminal.
Parameters
<name> description
BUS
Select software trigger.
IMMediate
Select immediate internal trigger.
EXTernal
Select hardware trigger from the rear-panel EXT TRIG terminal.
Query
command description
TRIGger:SOURce?
Query the trigger source.
Details
Refer to Trigger sources (on page 3-30) for information about sources.

Table of Contents

Other manuals for Keithley 2110-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110-100 and is the answer not in the manual?

Keithley 2110-100 Specifications

General IconGeneral
BrandKeithley
Model2110-100
CategoryMultimeter
LanguageEnglish

Related product manuals