5: Remote commands Model 2110 5½ Digit Multimeter
Reference Manual
5-42 2110-901-01 Rev. C/August 2013
[SENSe:]FRESistance:RANGe:AUTO <b>
This command turns autorange on or off for Ω4.
Parameters
<b> description
0 or OFF
Turn autorange off.
1 or ON
Turn autorange on.
Query
command description
[SENSe:]FRESistance:RANGe:AUTO?
Query Ω4 autorange state.
[SENSe:]FRESistance:RESolution <n>|MIN|MAX
Use this command to set a measured resolution for 4-wire (Ω4) measurements.
When you set the resolution, the instrument automatically updates the integration time.
Parameters
<n> description
(3.00e-07 to 3.00e-04) *
Ω
4 range
Select the resolution of the measurement.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command description
[SENSe:]FRESistance:RESolution?
Query the resolution setting.
[SENSe:]FRESistance:RESolution? MIN
Query the minimum resolution setting.
[SENSe:]FRESistance:RESolution? MAX
Query the maximum resolution setting.
Details
Refer to Integration time and resolution for DC and resistance measurements (on page 3-41) more information
about resolution.

Table of Contents

Other manuals for Keithley 2110-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110-100 and is the answer not in the manual?

Keithley 2110-100 Specifications

General IconGeneral
BrandKeithley
Model2110-100
CategoryMultimeter
LanguageEnglish

Related product manuals