3: Basic DMM operation Model 2110 5½ Digit Multimeter
3-40 2110-901-01 Rev. C/August 2013
Remote operation
Commands that use the <n> parameter can use the MINimum and MAXimum name parameters. MIN
selects the minimum value and MAX selects the maximum value.
Use the following commands to set a measurement range:
<function>:RANGe <n> Specify expected reading
<function> = VOLTage[:DC] <n> = -1000 to 1000 Volts DC
= VOLTage:AC = 0 to 750 Volts AC
= CURRent[:DC] = -10 to 10 Amps DC
= CURRent:AC = 0 to 10 Amps AC
= RESistance = 0 to 100e6 Ohms (2-wire)
= FRESistance = 0 to 100e6 Ohms (4-wire)
= FREQuency[:VOLT] = 0 to 750 Volts AC (voltage input)
= FREQuency:CURRent = 0 to 10 Amps AC (current input)
= PERiod[:VOLT] = 0 to 750 Volts AC (voltage input)
= PERiod:CURRent = 0 to 10 Amps AC (current input)
= CAPacitance = 0 to 10e-3 Farads
<function>:RANGe:AUTO <b> Enable or disable autorange
<function> = VOLTage[:DC] <b> = 0 or OFF Disable autorange
= VOLTage:AC = 1 or ON Enable autorange
= CURRent[:DC]
= CURRent:AC
= RESistance
= FRESistance
= FREQuency[:VOLT]
= FREQuency:CURRent
= PERiod[:VOLT]
= PERiod:CURRent
= CAPacitance