3: Basic DMM operation Model 2110 5½ Digit Multimeter
Reference Manual
3-46 2110-901-01 Rev. C/August 2013
Front-panel operation
Perform the following steps to set the aperture (integration time):
1. Select the frequency or period measurement function:
• To select frequency, press the FREQ key.
• To select period, press FREQ and then CONFIG. With PERIOD displayed, press ENTER.
1. Select the FREQ or PERIOD measurement function.
2. Press the CONFIG key, use the â—„ and â–º keys to display APERTURE, and then press ENTER.
3. Use the â—„ and â–º keys to display a aperture time and press ENTER.
Perform the following steps to set display resolution:
1. Select the frequency or period measurement function:
• To select frequency, press the FREQ key.
• To select period, press FREQ and then CONFIG. With PERIOD displayed, press ENTER.
1. Use the DIGITS key to set display resolution. Each key-press toggles the display between 4½
and 5½ digits.
Remote operation
Commands that use the <n> parameter can use the MINimum and MAXimum name parameters. MIN
selects the minimum value and MAX selects the maximum value.
Use the following commands to set the aperture time:
FREQuency:APERture <n> Set aperture time for FREQ
PERiod:APERture <n> Set aperture time for PERIOD
<n> = 0.01 0.01 s aperture
= 0.1 0.1 s aperture
= 1 1 s aperture

Table of Contents

Other manuals for Keithley 2110-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110-100 and is the answer not in the manual?

Keithley 2110-100 Specifications

General IconGeneral
BrandKeithley
Model2110-100
CategoryMultimeter
LanguageEnglish

Related product manuals