Section 5: SCPI command reference Model 2231A Triple-channel DC Power Supply Reference Manual
5-8 077100401/September 2014
This command is to set the voltage calibration points. P1, P2, P3 and P4 must be calibrated in numeric order.
Item Description
Group Calibration
Syntax CALibrate:VOLTage:LEVel <level>
Argument P1|P2|P3|P4
Returns None
Examples Cal:volt:lev P1
CALibrate:VOLTage[:DATA] <numeric value>}
This command is to return the actual output voltage value of the calibration point.
Item Description
Group Calibration
Syntax CALibrate:VOLTage[:DATA] <NRf>
Argument Voltage level
Returns <NRf>
Examples CAL:VOLT 30.0002V
CALibrate:CURRent:LEVel
This command is to set the current calibration points. P1, P2 must be calibrated in numeric order.
Item Description
Group Calibration
Syntax CALibrate:CURRent:LEVel <level>
Arguments P1|P2
Returns <NRf>
Examples Cal:curr P1
CALibrate:CURRent[:DATA] {<numeric value>}
This command is to return the actual output current value of the calibration point.
Item Description
Group Calibration
Syntax CALibration:CURRent [:DATA] <NRf>
Arguments Current level
Item Description
Returns None
Examples CAL:CURR 3.0002A